Chapter 7 Calibrate; Introduction To Calibration; Turbidity - Emerson Rosemount Clarity II T56 Reference Manual

Turbidmeter
Hide thumbs Also See for Rosemount Clarity II T56:
Table of Contents

Advertisement

Reference Manual
00809-0100-3565
7
Calibrate
7.1

Introduction to calibration

Calibration is the process of adjusting or standardizing the transmitter to a lab test or a
calibrated laboratory instrument or standardizing to some known reference.
Completing Quick Start upon first power-up enables live measurements but does not
ensure accurate readings in the lab or in process. Calibrate each attached sensor to ensure
accurate, repeatable readings.
7.2

Turbidity

This section describes how to calibrate the turbidity sensor against a user-prepared
standard as a two-point calibration with deionized water, against a 20 NTU user-prepared
standard as a single point calibration, and against a grab sample using a reference
turbidmeter.
To calibrate the turbidity sensor:
1. Press ENTER/MENU from the main screen to access the Calibration screen.
2. Select S1 Measurement or S2 Measurement and press ENTER/MENU.
3. Press INFO at any time to learn more about this procedure. When you press INFO, a
The following calibration routines are covered:
Slope Calibration: Slope calibration with pure water and a standard of known turbidity.
Rosemount T56
Figure 7-1: Calibration Screen
yellow screen appears with detailed instructions and information.
Calibrate
July 2020
45

Advertisement

Table of Contents
loading

Table of Contents