Maintenance
July 2020
6. Insert the replacement board in the sensor and push the socket on the replacement
7. Place the dessicant package in the sensor body.
8. Orient the flange so that the screw holes line up with the holes in the sensor body.
9. Place the sensor in the calibration cup and reconnect the cable.
10. Calibrate the sensor using either slope or standard calibration.
52
board into the mating pins in the sensor.
CAUTION
Equipment damage
Keep wires pushed away from lamp board when replacing the flange.
A. Wires and flange
B. Mating pins
Push the flange back on the sensor body and replace the screws.
Don't let wires push on lamp board. You may need to turn the flange a small
amount until the holes line up.
See
Calibrate slope
or
Calibrate against a
Failure to calibrate the sensor may reduce the life of the sensor.
Reference Manual
00809-0100-3565
standard. Do not use grab calibration.