Reference Manual
00809-0100-3565
6. Press ENTER/MENU.
7. To set relay logic to activate alarms at a high reading or a low reading, select High or
8. To enter setpoints for relays, enter the desired value for the process measurement
9. To set deadband as a measurement value, enter the change in the process value
10. To set the Normal alarm condition, select Open or Closed and press ENTER/MENU.
11. To set USP Safety, enter the percentage below the limit at which to activate the
5.7

Simulate relay action

To simulate relays:
Procedure
1. Press ENTER/MENU from the main screen to access the Program screen.
2. Select the Relay tab.
3. Select Simulate relay action.
Rosemount T56
Bleed and Feed
Water Meter
Display timer
Date and Time
Fault
None
Low and press ENTER/MENU.
or temperature at which to activate an alarm event and press ENTER/MENU.
needed after the delay deactivates to return to normal and press ENTER/MENU.
This prevents repeated alarm activation.
Program the Normal state to define the desired alarm default state to normally
open or normally closed upon power up.
alarm and press ENTER/MENU.
Note
USP Safety only appears if a contacting conductivity board is installed.
See
Figure
5-4.
You can manually set alarm relays in order to check devices, such as valves or
pumps.
Under the Alarms Settings menu, the following screen appears to allow manual
forced activation of alarm relays.