Fig. 2-12: Terminal Block X1 - Analog Signals And Relay Outputs 1-4 (Xsta) - Emerson Rosemount Analytical X-STREAM Enhanced Series Short Form Manual

Gas analyzers
Table of Contents

Advertisement

Short Form Manual
HASXEE-SFM-HS
05/2017
2.6.2 Installation - X-STREAM XEXF Field Housings
Analog signals
Relay outputs 1-4
Terminals for analog signals and relay outputs 1 - 4 are located on the outer left module (termi-
nal block X1).
Specification of analog signal outputs:
Specification of relay outputs:
Electrical specification:
Notes!
Consider
the installation notes in
the notes on installing cable glands on
Technical data and installation information for further available interfaces are given in the
X-STREAM Enhanced instruction manual. Connector layouts are part of the appendix of this
manual.
Serial Interface
*)

Fig. 2-12: Terminal Block X1 - Analog Signals and Relay Outputs 1-4 (XSTA)

Emerson Process Management GmbH & Co. OHG
4(0) – 20 mA; burden: R
Dry relay change-over contacts can be used as
NO or NC.
max. 30 VDC, 1 A, 30 W
2.7
Relay Outputs
**)
≤ 500 Ω
B
page 2-19.
Configuration of relay
**)
output terminals as per
standard factory setting
(NAMUR status signals).
NAMUR status signals are
automatically configured
Fail Safe!
Analog Outputs
X-STREAM XE
2-21