Appendix C: Specifications - Siemens CLINITEK Status Operator's Manual

Table of Contents

Advertisement

10
Appendices

Appendix C: Specifications

Appendix C: Specifications
Power Required
110V AC ± 20%, 45 - 65 Hz
(US only)
220V AC ± 20%, 45 - 65 Hz
(Europe only)
240V AC ± 20%, 45 - 65 Hz (UK
only)
100V - 240V AC ± 20%, 45 - 65 Hz
(with in-line lead)
Battery Powered Operation
Size 6 AA alkaline batteries
Dimensions
Depth - 272 mm (10.7 inches)
Width - 171 mm (6.7 inches)
Height - 158 mm (6.2 inches)
Weight
®
Clinitek Status
+ instrument only
(unpacked, without batteries or
power supply) – 1.66 kg (3.65 lb)
Ambient Operating Temperature
Range
18°C to 30°C (64°F to 86°F)
Ambient Operating Humidity
Range
18% to 80% Relative Humidity
(non-condensing)
10-8
Optimum Operating Temperature
Range
22°C to 26°C (72°F to 79°F)
Optimum Operating Humidity
Range
35% to 55% Relative Humidity
(non-condensing)
Optimum ranges insure that the
reagent results are optimized for
performance. At temperatures
under 22°C (72°F), urobilinogen
and leukocyte results may be
decreased, and at temperatures
above 26°C (79°F), increased.
Altitude: 2000 m (6562 ft)
Installation Category: II
Pollution Degree: 2
Instrument Memory
950 Patient test results
200 Patient details (Patient's Name
and/or Patient Identification)
Safety Standards
The Clinitek Status+ analyzer is
classed as a Class A computing
device in accordance with Part 15
of FCC Rules.