Siemens CLINITEK Status Operator's Manual page 115

Table of Contents

Advertisement

9
Cleaning and Maintenance
Cleaning
4
If the white calibration bar is
clean and unmarked, replace
the table into the analyzer by
holding the table at the end
opposite the white calibration
bar, with the white calibration
bar facing upwards. Push the
test table firmly but slowly, just
over half way into the analyzer.
Do not push the test table fully
into the analyzer as the test table
may become jammed and prevent
the use of the analyzer.
5
Replace the test table insert.
6
If the white calibration bar is
dirty or discolored, gently wipe
and clean it with a new cotton-
tipped stick or lint-free cloth
wetted with distilled water.
Care should be taken not to
scratch the white calibration bar.
Solvents of any kind must not be
used to clean the bar.
7
Allow the calibration bar to air
dry and then inspect the surface
for dust, foreign material,
scratches or scuffs. If the
calibration bar cannot be
cleaned or is still marked, obtain
a new test table.
8
Reinsert the test table as
described earlier in point 4.
9-5