Siemens CLINITEK Status Operator's Manual page 39

Table of Contents

Advertisement

4
Testing
Quick Tests
The Clinitek Status+ analyzer will
perform an automatic calibration each
time a test is run.
Be sure not to move or bump
the table while the instrument is
calibrating.
The Analyzing screen will be
displayed when the calibration has
been completed and the analysis of
the cassette has begun.
A timer will count down the time
remaining in analyzing the cassette
results.
The Clinitest hCG test results are
either negative, positive or borderline.
The analyzer takes approximately
5 minutes to confirm a negative
result. If the result is a clear positive,
the analyzer will report it sooner. If
the result is borderline, then you
should retest, with a new sample, in
48 to 72 hours. Please refer to the
Clinitest hCG cassette test package
insert for complete instructions for
use.