Mechanical Installation - HBM T10FM Mounting Instructions

Hide thumbs Also See for T10FM:
Table of Contents

Advertisement

10
SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182 E-MAIL:szss20@163.com
4

Mechanical installation

WARNING
Handle the torque flange carefully. The transducer might suffer perma-
nent damage from mechanical shock (dropping), chemical effects (e.g.
acids, solvents) or thermal effects (hot air, steam).
With alternating loads, you should use a screw locking device (medium)
to glue the rotor connection-screws into the counter thread to exclude a
pretension loss due to screw slackening.
An appropriate shaft flange enables the T10FM torque flanges to be mounted
directly. It is possible to mount a cardan shaft or relevant compensating ele-
ment directly on flange B (using an intermediate flange when required). Under
no circumstances must the permissible limits specified for bending moments,
transverse and longitudinal forces be exceeded. Due to the T10FM torque
flanges' high torsional stiffness, dynamic changes on the shaft run are minimi-
zed.
CAUTION
Do in any case check the effect on critical speeds and natural torsional
vibrations to avoid an overloading of the gaging flanges due to reso-
nance step-up.
NOTE
For regular operation do in any case observe the mounting dimensions
(see page 40).
NOTE
Even if the unit is installed correctly, the zero point adjustment made at
the factory can shift by approx. "100 Hz. If this value is exceeded, we
advise you to check the mounting conditions.
HBM
SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com
T10FM
B 23.T10FM.10 en

Hide quick links:

Advertisement

Table of Contents
loading