Settings - HBM T10FM Mounting Instructions

Hide thumbs Also See for T10FM:
Table of Contents

Advertisement

T10FM
SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182 E-MAIL:szss20@163.com
NOTE
The gaging flange should not be under load when the calibration signal
is being measured, since the calibration signal is mixed additively.
CAUTION
To maintain measurement accuracy, the calibration signal should be
connected for no more than 5 minutes. A similar period is then needed
as a cooling phase before triggering the calibration signal again.
Applying a 5 V separated extra-low voltage to pins 6 (+) and 7( ) on plug 1 or
3, triggers the calibration signal.
The nominal (rated) voltage for triggering the calibration signal is 5 V (trigge-
red when U>2.7 V). The trigger voltage is galvanically isolated from the supply
voltage and the measurement voltage. The maximum permissible voltage is
30 V
At voltages lower than 0.7 V, the gaging flange is in measuring mode.
.
Current consumption at nominal (rated) voltage is approx. 2 mA and at maxi-
mum voltage is approx. 22 mA.
NOTE
In the case of HBM system solutions, the measuring amplifier triggers
the calibration signal.
7

Settings

NOTE
You will find a table containing all the relevant switch positions on the
back of the stator cover. Changes to the factory settings should be no-
ted here using a waterproof felt-tip pen.
B 23.T10FM.10 en
SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com
29
HBM