Maintenance and service
9.3

Repair electronics housing

Use the following procedures to remove damaged components from the transmitter
electronics housing and to install new replacement parts.
Disassemble the unit only as needed to replace damaged components. Use the assembly
procedures that apply to install replacement parts and reassemble the unit.
9.3.1

Electronics housing disassembly

Remove cover
Complete the following steps to remove the cover from the electronics housing on the
OCX 8800.
1.
248
The reading should be from 10 to 13 inches, Water Column. Locate and correct leaks
if the reading is less than 10 inches WC.
See
Figure
9-22. Loosen screw (L) and slide locking clip (K) away from cover (A).
Retighten screw (L).