Table 7-3:
Descriptions of PlantWeb Alerts
Alerts
Alarms
Sensor Degraded
Transducer block:
O
Cell Impe-
2
dance High
Reference Manual
(continued)
What does the
AMS tab
alert indicate?
Sensor Alerts
This alert is active
when the oxygen
sensor impedance
indicates that the
cell is beyond its
useful life.
Foundation Fieldbus
Recommended
action
Description
also cause this
alarm. Refer
to
Section 8.2
for details.
Replace the oxy-
Oxygen cells de-
gen cell.
grade over time
due to aging and
corrosion. An in-
creasing cell re-
sistance is a good
indicator of re-
duced cell per-
formance. As the
cell impedance in-
creases, the cell
output falls off,
and response
time increases.
Calibrating the in-
strument com-
pensates for the
increased cell re-
sistance up to sev-
eral hundred
ohms, beyond
which the cell is
no longer func-
tional. Diagnosis
must be done at
the analyzer. Re-
fer to
Section 8.2
for details.
103