Emerson Rosemount OCX 8800 Reference Manual page 251

Oxygen and combustibles transmitter
Hide thumbs Also See for Rosemount OCX 8800:
Table of Contents

Advertisement

5.
6.
Reference Manual
Figure 9-18:
COe sensor holder alignment
A. Sensor holder flat
B. Matchmark
C. Straight edge
D. Matchmark
CAUTION!
EQUIPMENT DAMAGE
The heater insulator prevents current leakage between the band heater and the sensor
holder. Failure to properly install the insulator may cause the device to trip a ground
fault interrupt circuit.
Wrap heater insulator (K) around sensor holder (J). Make sure the insulator joint lines
up with the band gap of the COe band heater (M).
Slide the COe band heater (M,
Figure
9-15) up onto sensor holder (J).
Maintenance and service
241