Terminal Assignment Connector "E" (Rs485 - Modbus); Terminal Assignment Connector "A - B" (Alarm Relay) - Beko METPOINT BDL compact Installation And Operating Manual

Hide thumbs Also See for METPOINT BDL compact:
Table of Contents

Advertisement

Installation and operating manual

9.4. Terminal assignment connector "E" (RS485 - Modbus)

9.5. Terminal assignment connector "A – B" (alarm relay)
The alarm outputs are designed as floating switchover contacts. The alarm signal can be forwarded to a control desk, etc. 
through the floating contacts. The connectors for the alarm contacts are labelled as "A" and "B".
DANGER!
Mains voltage
When wiring the electric power supply line, ensure that the double or reinforced insulation between
the electric circuits and the secondary circuit remains intact.
NOTICE!
Mains voltage
The additional insulation must be suitable for a test voltage of 1500 VAC. The insulation must be at
least 0.4 mm thick
e.g. insulation hose, type BIS 85 from Bierther GmbH.
The additional insulation of the power cables (mains connection, alarm and signalling relays) can be implemented as follows:
1
- Terminals (plug-type connectors)
2
- Flexible insulating tubing for the power cables
3
- Connecting cable
20
1
2
3
3
3
4
5
6
1
2
1
2
EN
METPOINT
BDL compact
®

Advertisement

Table of Contents
loading

Table of Contents