The surface of the Connector Hose can be cleaned using a soft cloth and one of the
following:
Mild detergent
Steri-Fab
®
Antifect
FF
®
Mikrozid
Sensitive Wipes
®
70% Isopropyl alcohol
We do not recommend the use of quaternary ammonium (such as Virex
DO NOT USE:
Phenolic-based disinfectants (such as Amphyl
setting).
Any solvent based cleaners. Doing so will damage the plastics and will void your
warranty.
Abrasive materials. Doing so will damage the plastics and will void your warranty.
Any petroleum based lubricants. Doing so will damage the o-rings and will void your
warranty. If lubrication is required, the use of silicon spray is recommended.
Gently remove Heat Exchanger from the Sleeve. Hand wash the Sleeve in cold water,
using a mild detergent or antibacterial soap. Do not use fabric softener. Hang to dry.
If needed, the external surface of the Heat Exchanger may be cleaned by wiping down with
commercial non-bleach cleaning wipes or hand washed using a very small amount of mild
detergent or antibacterial soap. DO NOT MACHINE WASH.
Refer to the Wrap Use Guide accompanying individual Wraps for more information.
The Carry Bag should be cleaned using a soft cloth or brush and a mild detergent. Febreze
or the equivalent can be used on the Bag if desired. If the Carry Bag has a biological material
on the surface, Steri-Fab
Be sure to test any product on a small portion of the Bag to ensure that it will not cause
damage.
Note: To operate the GRPro 2.1 System, you do not need to remove it from the Carry Bag.
Simply unzip the Bag's main compartment and end panel. Fill the reservoir with ice and
water. Attach the Connector Hose and the AC Adapter to the end panel of the Control Unit
and plug the AC Adapter into an electrical outlet.
CONNECTOR HOSE
WRAPS
CARRY BAG
may be used to decontaminate those surfaces.
®
– typically only found in a clinical use
®
) or Cavicide.
®
®
®
EN-13

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GRPro 2.1 CONTROL UNIT and is the answer not in the manual?

Questions and answers

Salmi Mustapha
January 23, 2025

Bjr j'ai un Game ready il marque erreur 01puis 02 puis 06

User image 679281a474a89

This manual is also suitable for:

550550-03550550-03-rn550550-53Grpro 2.1 control unit

Table of Contents

Save PDF