testo 6681
56
Enter reference value for 1-point adjustment
1
In Measuring Mode, press SET, select Main Menu Adjust using or
and confirm selection with SET.
Reference value % RH is displayed.
2
Press SET, edit value: Scroll one digit to the right using , increase
value of digit by 1 using . Confirm with SET or cancel entry with
ESC.
3
Continue to Reference value temp with or .
4
Press SET and Reference value °C is displayed.
5
Press SET, edit value: Scroll one digit to the right using , increase
value of digit by 1 using . Confirm with SET or abort entry with ESC.
6
Continue to Reference value °F with or .
7
Press SET, edit value: Scroll one digit to the right using , increase
value of digit by 1 using . Confirm with SET or abort entry with ESC.
8
Continue to Analog Adj. Ch. 1 with or .
9
Continue with the adjustment of the analog outputs (see below, step 2)
or press ESC to return to Main Menu Adjustment.
10 Continue to Main Menu Reset using or or return to Measuring
Mode by pressing ESC.
Performing the 2-Points-Adjustment for 20% RH and 80 % RH
1.
In the Measuring Mode press SET, select main menu Adjust with or
and confirm selection with SET.
2.
Continue to 2Pt Adjust 20/80% with or .
3.
Press SET.
4.
Continue to Adj. Point 20% or Adj. Point 80% with or
5.
Confirm selection with SET.
The red LED ADJ flashes
In the display 2Pt Adjust 20/80%, 1-point adjustment, Probe reset
appears consecutively.
The adjusted measurement values appear after probe reset.
- 1.4 Operation
Please also refer to chapter 1.3.3.2.