For the effectiveness of the self-adjustment the following prerequisites are
decisive over the duration of the self-adjustment.
The process temperature should not vary by more than 0.5K
The dew point temperature should remain stable, as far as possible
The process pressure should not vary excessively
All adjustment processes are saved in the history of the transmitter, see
Chapter 3.3.5 Transmitter history.
testo 6615:
testo 6615:
probe
Fühler-
behaviour
verhalten
Analog
Analog-
output °Ctd/
Ausgang
°Ftd
°Ctpd/
°Ftpd
- 1.3 Commissioning
If these prerequisites cannot be fulfilled, the values obtained
during the last successful self-adjustment will be maintained.
Parameter
Parameter
"Duration self-adjustment time"
„Dauer Heizphase"
"Self-adjustment cycle time"
„Zykluszeit Selbstabgleich"