TESTO 6681 Instruction Manual page 141

Humidity transmitters
Hide thumbs Also See for 6681:
Table of Contents

Advertisement

Column
Explanation
Actual value
1-point adjustment: If no changes were performed, no
before
value is displayed.
adjustment
Offset from
1-point adjustment: Value before the adjustment.
Offset to
1-point adjustment: Value after the adjustment.
Offset
2-point adjustment: Difference between target and actual
value reported by instrument.
Channel
Analog adjustment: Channel 1 to n.
Specification
Analog adjustment: Actual value.
Reading
Analog adjustment: Reference value entered.
Offset
Analog adjustment: Deviation at time of adjustment.
P2A software
- 3.3 Using the software 141