Procedure - Atmos iQam Operating Instructions Manual

Endoscopic camera
Hide thumbs Also See for iQam:
Table of Contents

Advertisement

5.3.3

Procedure

Pretreatment at the site
of use
Pre-clean
Dismantle
Manual cleaning and
disinfection
Checking and maintain-
ing
Assembly
Function check
Packaging
Sterilizing
20
Reprocessing
Not necessary
Not necessary
Not necessary
1.
Clean the device evenly with a suitable cloth
and a suitable detergent under running
water.
2.
Immerse the device in the cleaning solution
(follow manufacturer's instructions) and clean
it in the solution evenly with a suitable cloth.
3.
Disinfect the device evenly with a suitable
cloth and a suitable disinfectant. Observe
instructions regarding contact time.
1.
Check wether reprocessing was successful
with a suitable light magnifier.
2.
If reprocessing was unsuccessful, the proce-
dure must be repeated.
Not necessary
Other: coupler shoe function
Not necessary
Not necessary