Atmos Record 55 DDS Operating Instructions Manual

Atmos Record 55 DDS Operating Instructions Manual

Hide thumbs Also See for Record 55 DDS:

Advertisement

Quick Links

English
ATMOS MedizinTechnik
GmbH & Co. KG
ATMOS
Record 55 DDS
Ludwig-Kegel-Str. 16
79853 Lenzkirch / Germany
MedizinTechnik
Tel. +49 (0) 7653 / 6 89-0
Fax +49 (0) 7653 / 6 89-190
444.0910.B
444.0930.B
444.0940.B
2013-09 Index: 20
atmos@atmosmed.de
www.atmosmed.de

Advertisement

Table of Contents
loading

Summary of Contents for Atmos Record 55 DDS

  • Page 1 MedizinTechnik English ATMOS Record 55 DDS 444.0910.B 444.0930.B 444.0940.B 2013-09 Index: 20 ATMOS MedizinTechnik Ludwig-Kegel-Str. 16 Tel. +49 (0) 7653 / 6 89-0 atmos@atmosmed.de GmbH & Co. KG 79853 Lenzkirch / Germany Fax +49 (0) 7653 / 6 89-190 www.atmosmed.de...
  • Page 2: Table Of Contents

    4.15.2 Foot regulator, Art.No. 443.0770.0, with the ATMOS Record 55 DDS ......15 4.15.3 Foot regulator, Art.No. 443.0770.0, with the ATMOS Record 55 DDS DDS Lipo ..15 4.15.4 Mobility of the ATMOS Record 55 DDS ....15 Cleaning General information on cleaning and disinfection ............16...
  • Page 3: Introduction

    • ATMOS Record 55 DDS Lipo REF 444.0920.0 on how to operate the Record 55 DDS safely, correctly and effectively. The operating instructions are intended for new as well as for variants, consisting of above mentioned basic operating personnel and also should be con-tinued to be units in connection with following options: used as a reference manual.
  • Page 4: Intended Use

    1.0 Introduction Intended use  The ATMOS Record 55 DDS is a mobile surgical suction unit for medical application. It is especially intended for aspiration and collection of secretions, body fl uids, tissue particles and rinsing solutions. Due to its high performance values, it is also most suitable for the subcutaneous lipo- suction.
  • Page 5: Explanation Of Symbols

    1.0 Introduction Explanation of symbols Attention, refer to operating instructions ! Fuse Potential equalization Type BF equipment Alternating current Protection against penetration of IPX1 damaging humidity (drop water) Foot switch Class AP, for use in explosion-hazardous areas...
  • Page 6: For Your Safety

    2.0 For your safety  The design of the ATMOS Record 55 DDS fulfi lls the re- quirements of IEC 601/EN 60601 and of protection class I. The device must only be connected to a properly installed socket with non-fused earthed wire.
  • Page 7  The ATMOS Record 55 DDS fully complies with the electro- magnetic immunity requirements of standard IEC 601-1-2 / EN 60601-1-2 "Electromagnetic compatibility - Medical Electrical Equipment".
  • Page 8: Figures

      Fig. 1. ATMOS Record 55 DDS, Front View Fig. 2. ATMOS Record 55 DDS Lipo, Front View (illustration with canister set 2 x 3 l DDS) (illustration with canister set 2 x 5 l DDS)   Connection for foot switch (option) Connection for foot switch (option) ...
  • Page 9 3.0 Installation and start-up          Operating Controls and Indicators ATMOS Record 55 DDS, Rear View Fig. 3. Fig. 4.   ON/OFF switch Connection for potential equalisation  (ON = continuous operation) Fuse ...
  • Page 10: Operation

    Replace the DDS bacterial fi lter at least once a day. Use  Fig. 6. Bacterial fi lter only original ATMOS bacterial fi lters!  DDS canister handle Never operate the unit without the DDS bacterial fi lter/ oversuction stop as otherwise the warranty will expi- ...
  • Page 11: Attach / Remove Dds Secretion Canister Lid

    4.0 Operation Attach / remove DDS secretion canister lid  With the DDS secretion canister on a fi rm surface, position the lid horizontally on top (the lid may not be twisted!)  Press down lightly onto the secretion canister using both hands until limit is reached.
  • Page 12: Secure Dds Secretion Canister

    Twist slightly in the same manner when removing. Fig. 13. 4.10 Connect hose 4.11 Attaching the tumescence infi ltration pump Fig. 14. ATMOS S 2001 TI  See assembly instructions. Assembly instructions are enclosed to the support for the  tumescence infi ltration pump.
  • Page 13: Checking The Supply Voltage

    15).   For surgical use, we recommend to connect pin (  fi g. 15) of the ATMOS Record 55 DDS to the room's  potential equalisation system. The ATMOS Record 55 DDS is then ready for operation. ...
  • Page 14: Suction

    Damage on device at higher weight!  The DDS changeover docking station for 2 secretion ca- nisters with changeover switch and direct docking may be used on the ATMOS Record 55 DDS.   The changeover lever ( , fi g. 18) serves to switch the vacuum to the secretion canister towards which the lever Fig.
  • Page 15: Options

     Hoses and cables must be secured on the back before the unit is moved. Stand behind the ATMOS Record 55 DDS and push it with  one or both hands on the level of the display and operating panel.
  • Page 16: Cleaning

    5.0 Cleaning General information on cleaning and disinfection  For disinfection, you may use all surface and instrument disinfectants listed in chapter 5.4 / 5.5.  A number of disinfection agents may cause discoloration at the secretion canister etc., however this has no effect upon the parts's function.
  • Page 17: Cleaning And Disinfecting The Unit Surface

    5.0 Cleaning Cleaning and disinfecting the unit surface  Always disconnect the device from the power line, before cleaning and disinfecting the surface.  Wipe the surface clean with a cloth soaked in a cleaning solution or disinfectant. Liquids must not enter the device. All of the cleaning solutions and disinfectants listed below can be used.
  • Page 18: Maintenance

    Airtight packing  Please enclose a detailed error description. Warranty ATMOS cannot guarantee an error-free function nor can ATMOS be held liable for damage to people or goods if  non-original ATMOS parts are used,  the information in these operating instructions are disregarded, ...
  • Page 19: Spare Parts And Accessories

    7.0 Trouble-shooting The ATMOS Record 55 DDS was subjected to a thorough quality control before shipment. If there is, nevertheless, some mal- function, you possibly might solve this problem yourselves if you observe the following instructions. Problem Possible causes Remedy ...
  • Page 20: Accessories

    8.0 Spare parts and accessories Accessories 8.2.2 Facilities to simplify the handling Description Foot regulator set ..................443.0770.0 Foot switch ....................443.0755.0 Tray with rail adaptor ................... 443.0790.0 Hose holder on canister................340.0066.0 Catheter quiver for fl ex. catheters ............... 444.0140.0 Catheter quiver with holder for rail system (for catheter storing) ..................
  • Page 21 8.0 Spare parts and accessories 8.2.4 Gynaecology Description Suction curette, with auxiliary air vent, external Ø 6 mm......401.0529.0 Suction curette, with auxiliary air vent, external Ø 8 mm......401.0530.0 Suction curette, with auxiliary air vent, external Ø 10 mm......401.0531.0 Suction curette, with auxiliary air vent, external Ø...
  • Page 22: Technical Specifi Cations

    9.0 Technical Specifi cations Technical Specifi cations ATMOS Record 55 DDS Air fl ow rate of pump 55 ± 3 l/min. Max. vacuum at sea level -98 kPa (-980 mbar or -735 mmHg)* Vacuum readout -1...0 bar (± 25 mbar)
  • Page 23 9.0 Technical Specifi cations Technical Specifi cations ATMOS Record 55 DDS Lipo  As with the ATMOS Record 55 DDS (Chapter 9.1) but REF 444.0920.0.  Technical Specifi cations for the tumescence infi ltration pump ATMOS S 2001 TI are contained in the respective...
  • Page 24: Disposal

     The material of the housing can be recycled completely.  The component parts of the ATMOS Record 55 DDS must be disposed off correctly and the materials are to be separated carefully. This document is subject to change without notice.
  • Page 25: Notes On Emc

    11.1 Guidelines and Manufacturer´s Declaration - Emissions The ATMOS Record 55 DDS is intended for use in the electromagnetic environment specifi ed below. The customer or user of the ATMOS Record 55 DDS should ensure that it is used in such an environment.
  • Page 26 1.3 Guidelines and Manufacturer´s Declaration - Immunity The ATMOS Record 55 DDS is intended for use in the electromagnetic environment specifi ed below. The customer or user of the ATMOS Record 55 DDS should ensure that it is used in such an environment.
  • Page 27 To determine the electromagnetic environment in regard to stationary transmitters, a study of the location is to be considered. If the measured fi eld strength at the location where the ATMOS Record 55 DDS is used exceeds the above compliance level, the ATMOS Record 55 DDS is to be observed to verify the intended use.
  • Page 29 Any and all present or future government charges, clusive remedy will be the repair or, at ATMOS’s sole levies, or taxes on the products shall be added to option, the replacement of the defective product. At...

This manual is also suitable for:

Atmos record 55 dds lipo444.0910.0444.0920.0