Table of contents Introduction.......................3 Notices on the operating instructions ..............3 Explanation of pictures and symbols ..............4 Intended use .......................5 Function ........................6 Intended users ......................6 Scope of delivery ......................6 Transport and storage ....................6 Notices for your safety ..................7 General safety information ..................7 Danger for users, patients and third parties ............7 Avoid damage to the device..................8 Setting up and starting up ................9...
The Declaration of Conformity and our general standard terms and conditions can be viewed on our website at www.atmosmed.com. The quality management system at ATMOS has been certified according to international standard EN ISO 13485. These operating instructions are valid for the following devices: •...
1.2 Explanation of pictures and symbols In the operating instructions DANGER Warning of a danger that results in immediate fatal or serious injury. Observe the necessary measures. WARNING Warning of a danger that can cause fatal or serious injury. Observe the necessary measures.
For indoor use only Professional disposal Equipotential bonding Alternating current Earth Maximum load Max. Do not transport any weights or persons on the chair (Mobile/Complete versions). 1.3 Intended use Intended use • As a treatment chair for treatment and examination on patients who are seated or lying down •...
If damage occurs during transport: 1. Document and report the transport damage. 2. Send the device to ATMOS, see Chapter “3. Check whether all the fixing screws are fully tightened.” on page 18. Ambient conditions for transport and storage: •...
If any signs are missing from your device or no longer legible, contact ATMOS and replace the signs in question. 2.2 Danger for users, patients and third parties Electric shock due to unsuitable mains connection, incorrect handling of the product or damage to product components.
Observe the specifications regarding period tests in chapter “6.1 Period tests” on page 18. • Only use original accessories and original spare parts from ATMOS. • Assembly, new settings, modifications, extensions and repairs may only be carried out by persons who are authorised by ATMOS.
3 Setting up and starting up 3.1 Device overview Front view Headrest Backrest Control panel (Electrical/ Complete versions) Arm rest Backrest release lever (Basic/ Mobile versions) Equipotential bonding connection Release lever for top section of chair Roller foot pedal (Mobile/ Basic Mobile Complete versions)
Control panel (Electrical/Complete versions) button ackrest Back button ackrest forward button hair up button hair down 3.2 Connecting the device 1. Clean and disinfect the product prior to first use. 2. Connect the equipotential bonding if required. The equipotential bonding must be connected if required by the spatial environment in which the product is going to be used.
4 Operation 4.1 Ambient conditions during operation The product is intended exclusively for use in buildings and under normal ambient conditions. Do not subject the upholstery to high temperatures. • Temperature: +10...+35°C • Relative humidity: 20...95% without condensation • Air pressure: 700...1060 hPa The device can only be disconnected from the power supply by pulling out the power plug.
4.3.2 Adjusting the foot support Unfolding the foot support 1. Pull the handle and fold the foot support down. Folding the foot support away 1. Fold the foot support up. 4.3.3 Adjusting the backrest Mechanical 1. Push the backrest release lever down, pushing or pulling on the backrest at the same time.
Height 1. Undo the headrest locking screw (height) move the headrest to the required position. 2. Tighten the headrest locking screw (height) » The headrest is fixed in place at the current height. 4.3.5 Adjusting the seat height 1. Press the corresponding foot switch ...
4.4 Adjusting the shock position 1. Adjust the angle of the backrest up to a fully horizontal position. 4.5 Moving the Patient Chair (Mobile/Complete versions) ATTENTION Damage caused by moving the patient chair in contravention of intended use. There is a risk of damage to the patient chair and the power cable. •...
4.6 Activating and deactivating the operation lock Activate the operation lock if you need to leave your patients alone, or if there is a risk of buttons and foot switches being operated unintentionally. Never leave elderly persons, children, injured persons, persons with disabilities or other patients who require assistance alone on the patient chair.
5 Cleaning and Disinfection We recommend that you always document all maintenance work and part replacements in writing. It is the responsibility of the user to ensure that the required results for cleaning and disinfection are adhered to. Validation and routine monitoring of the procedure will generally be necessary.
5.3 Recommended disinfectants ATTENTION Unsuitable cleaning agents and disinfectants. Risk of damage to the device surface, corrosion damage and stress cracks. • Do not use cleaning agents or disinfectants that contain the following ingredients: • Irritants • Substances that may damage plastics or steel components or dissolve lubricants. ATTENTION Unsuitable cleaning agents and disinfectants.
ATMOS recommends: Work should be carried out by an authorised ATMOS service partner. This ensures that the repairs and testing are carried out professionally, that original spare parts are used and that warranty claims remain unaffected.
Page 19
4. The device must be well padded and packed in suitable packaging. 5. Place the QD 434 “Delivery complaint/return shipment” form and the corresponding decontamination certificate in an envelope. 6. Affix the envelope to the outside of the package. 7. Send the product to ATMOS or your dealer. Maintenance and service...
» The system has been reset. Inexplicable noises Motor defective or worn 1. Disconnect the device from the detected. mains power supply. The motor is 2. Contact an authorised ATMOS becoming hot. service partner. 20 Troubleshooting...
1. Clean and disinfect the device. 2. In Germany: Send the product in to ATMOS or your specialist dealer. They will dispose of the device professionally. 3. In other countries: Dispose of the product professionally and according to the country-specific laws and regulations.
9 Technical data Voltage 230 V~ ± 10%; 50/60 Hz Current consumption Max. 2 A (230 V) Power consumption Max. 460 W Fuses T 6.3 A (for 230 V~, 50/60 Hz) Seat height 58 cm (rigid seat upholstery) Max. lifting capacity 200 kg (safe working load) Vertical speed Max. 18 mm/s Height of backrest Approx.
Classification according to Appendix Class 1 IX, EC Directive 93/42/EEC In accordance with Regulations 1 & 12, Section CE marking UMDNS code 10-794 ID No. (REF) 535.5200.0 Basic 535.5300.0 Electrical 535.5400.0 Mobile 535.5500.0 Complete 10 Notices on EMC As soon as the patient chair is connected to the mains power supply and highly sensitive measuring equipment is in use at the same time, it is impossible to guarantee that the patient chair is not electro-magnetically affecting the measurement results.