Exit From Head Calibration - Fluke Calibration 9500B Calibration Manual

High performance oscilloscope calibrator
Table of Contents

Advertisement

9500B
Calibration Manual

Exit from Head Calibration

When the necessary calibration operations are complete, return to the Calibration screen
that contains the Exit and Store Head Cal softkeys. Selecting Exit produces the reminder
message:
Select Exit to abandon head cal data.
To store head cal data, select the Store Head Cal softkey, which produces a new screen
labelled:
Follow the on-screen instructions to:
Below the displayed dates is a message that reads:
followed by a line that identifies the heads e.g. CH1, CH4. If there is no cal data to store,
or if all head cal data storage is complete, the NONE indicator appears.
Selecting the Store softkey displays the message:
The save operation may take up to 30 seconds to complete. To save any remaining
unsaved head data, use the Channel Select softkey to change to the appropriate head, then
select Store. Repeat for each unsaved head until the NONE indicator is displayed.
Select Exit to step back through the menus to the required instrument setting.
144
WARNING:
Head Calibration: CHx
Select the warning period before recalibration is due
Modify the cal due date
These heads have unsaved data
Saving cal factors to head x
Cal data not stored:
Exit again to abandon data
(where x is the head number)
(where x is the head number)