Isite Configuration - Cardinal 225D SmartCell Installation And Technical Manual

Hide thumbs Also See for 225D SmartCell:
Table of Contents

Advertisement

The 225D will periodically send indicator, scale, and cell data to the Cardinal iSite Webserver
for diagnostic logging. This information will be used by the server to determine if there are
problems with the scale(s) that need to be addressed.
To access the ISITE IP CONFIG menu:
Press SHIFT + RED_KEY to enter SETUP/REVIEW.
1.
Press ENTER once and DOWN twice to navigate to SETUP MENU #3.
2.
Select #9. ISITE IP CONFIG.
3.
The SO# of the scale is used to match up the scale to the correct iSite dealer account. In many
cases, DHCP may be used in which case setup is quite simple:
SO# = XXXXXX
1.
DHCP = YES
2.
If a static IP address is required (such as to address firewall issues), then set DHCP = NO,
and prompting will appear to manually set addresses:
SO# = XXXXXX
1.
DHCP = NO
2.
IP = XXX.XXX.XXX.XXX
3.
SUBNET = XXX.XXX.XXX.XXX
4.
GATEWAY = XXX.XXX.XXX.XXX
5.
To confirm that iSite is working or to diagnose any errors in the connection, there is an iSite
status page in the diagnostics menu. Some of the information is quite technical, but is present
in case of a more complicated problem.
To check status of iSite connection:
From the main weight screen go to the diagnostics menu by pressing SHIFT +
1.
CELL_DIAG.
Navigate with PREVIOUS/NEXT soft keys to page titled "ISITE STATUS OF LAST
2.
CONNECTION".

iSITE CONFIGURATION

8200-0753-0M Rev A  225D Weight Digital Indicator
31

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents