Indicator Connection
The homerun cable should be routed through the metallic gland connector installed in 225D
rear panel (lower right).
I/O
(Serial,
Isolated
Inputs or
Outputs)
Fiber
Optic
Cable
1. Remove the 14 acorn nuts securing the rear panel assembly to the main housing.
2. Lift the rear panel from the main housing, taking care not to stretch the cable and wires
between the panel and main housing. Lay the rear panel on the workbench next to the
indicator.
IMPORTANT: You may need to loosen the gland connectors for the I/O cables to
allow enough slack in the cable and wires to avoid stretching them.
3. Loosen and remove the metal gland connector nut and remove the plastic insert.
4. Route the homerun cable through the nut and plastic insert and into the enclosure.
5. With the homerun cable routed into the enclosure, remove approximately 2 1/4 inches of
the cables' outer braided jacket, exposing the inner insulating jacket.
4
AC Power
100-240 VAC
0.4 Amp Max.
ATTENTION!
OBSERVE PRECAUTIONS FOR HANDLING
ELECTROSTATIC SENSITIVE DEVICES
8200-0753-0M Rev A 225D Weight Digital Indicator
Optional
2XX-USBA
Connectors
I/O
(Serial,
Isolated
Inputs or
Outputs)