LOAD CELL X HAS BEEN REPLACED
NO WEIGHT
-------
LOAD CELL 2 HAS BEEN REPLACED.
SCALE NOT LEGAL FOR TRADE UNTIL
CALIBRATED.
PRESS ENTER TO CONTINUE
Probable Cause
A load cell has been replaced. Replaced load cells are automatically detected as long as
MORE LOAD CELLS THAN EXPECTED
NO WEIGHT
-------
MORE LOAD CELLS THAN EXPECTED
Probable Cause
There are more load cells
than expected connected to
the 225D. For example, the
indicator number of load cells
is set to 8 but the controller
detects there are actually 12
load cells.
only one load cell was replaced. Once a load cell has
been replaced, the 225D will auto detect the new load
cell, display the message that the scale is not Legal-For-
Trade, and will need to be calibrated.
ID
COUNT
Confirm and configure the number of load cells the scale
should have.