7.2 Advanced Settings (
Setting method
Setting item
Eco mode (ECO)

7.2.9 Alarm Settings (

This function allows you to specify whether to detect alarms and to set alarm detection
conditions.
Alarms are divided into two types (measurement alarm 1 and measurement alarm 2).
Measurement alarm 1 occurs when the sensor head becomes unable to make measurements.
This alarm occurs when an edge exceeding the measurement capability occurs or when the
intensity of received light is saturated due to ambient light.
Measurement alarm 2 occurs when the sensor head becomes dirty, when the number of
connected controllers is changed, or when the specified measurement direction differs from the
insertion direction of the detected object. (Each function can be set to ON or OFF.) Digital
measured values are output directly as measured values. However, they are reflected in
judgment output.
Alarm type
Measurement
alarm 1
Measurement
alarm 2
7-42
)
Eco OFF (OFF)
Eco ON (ON)
)
Description
This alarm occurs when the intensity of received
light is saturated due to ambient light or when an
edge exceeding the measurement capability
occurs.
This alarm occurs when the sensor head becomes
dirty, when the specified measurement direction
differs from the insertion direction of the detected
object.
Setting range
Measured
value
Previous value
or
ALARM
(optional)
Measured
value
Default value
OFF
Judgment output
Judgment output
based on the previous
value
or
alarm output
(optional)
Alarm output
WUME-HGT-6