h.
Same as b. above.
i.
Repeat CHECK for CH 4.
j.
Set CH 3 and CH 4 VOLTS/DIV to 0.5 V.
k.
Set calibration generator to Std Ampl output, 2 V.
.
CHECK—the signal amplitude is between 3.92 and 4.08 divisions.
l
m. Set:
VERTICAL MODE
CH 3 VOLTS/DIV
n.
Move the test signal to the CH 3 input.
o.
Repeat CHECK procedure for CH 3.
p.
Disconnect the test setup from the 2245A.
11.

ADD Mode and CH 2 INVERT Deflection Accuracy

a.
Set:
VERTICAL MODE
CH 1 and CH 2 VOLTS/DIV
CH 1 and CH 2 Input COUPLING
b.
Connect calibration generator Std Ampl output to the CH 1 and CH 2
inputs via a 50 A precision BNC coaxial cable and a BNC dual-input
coupler.
c.
Set the calibration generator to Std Ampl output, 0.2 V.
d.
Position the ADD signal to the center of the crt graticule with the CH 1
and CH 2 POSITION controls.
e.
CHECK—That the ADD signal amplitude is between 3.92 and 4.08 divi­
sions.
f.
Set CH 2 INVERT On.
g.
CHECK—That the ADD signal amplitude is 0.08 division (less than half
a minor graticule division) or less excluding trace width (sweep will free
run).
h.
Disconnect the test setup from the 2245A.
2245A Operators
Performance Check Procedure
CH 3 (CH 4 off)
0.5 V
ADD (all others off)
0.1 V
DC
7-17