Performance Check Procedure
c.
Connect calibration generator (PG 506) fast rise, rising-edge signal to
the CH 1 and CH 2 inputs via a 50 O precision BNC coaxial cable, a
50 f l BNC termination, and a BNC dual-input coupler.
d.
Connect calibration generator trig out signal to the CH 3 input via a
50 f l BNC coaxial cable and a 50
e.
Set the calibration generator to Fast Rise and adjust Pulse Amplitude
for five divisions of signal amplitude at 1 MHz.
f.
Position the rising edges of the superimposed waveforms horizontally to
the center vertical graticule line.
It may be necessary to readjust the trigger level to display the
rising-edge signal In the Delay Match steps. This can be done by
pressing the upper Trigger MODE button to cause the trigger
level to be automatically readjusted.
g.
Set X I0 MAG On (for 2 ns/div sweep speed).
h.
CHECK—that the leading edges of the two waveforms have less than
0.1 horizontal division separation at the center graticule line excluding
trace width.
14. CH 1 to CH 4 Signal Delay Match
a.
Set VERTICAL MODE to CH 1 and CH 4 (CH 2 o ff).
b.
Move the CH 2 signal to the CH 4 Input connector.
c.
Superimpose the CH 4 waveform on the CH 1 waveform.
d.
CHECK—that the leading edges of the two waveforms have less than
0.1 horizontal division separation at the center graticule line excluding
trace width.
15. CH 3 to CH 4 Signal Delay Match
a.
Set:
VERTICAL MODE
TRIGGER SOURCE
b.
Move the CH 3 signal to the CH 2 input and the CH 1 trigger signal to
the CH 3 input.
7-20
Cl
BNC termination.
NOTE
CH 3 and CH 4 (CH 1 off)