Performance Check Procedure
Equipm ent Required (See Table 7-1)
Calibration generator
Test oscilloscope
Level sine-wave generator
1. Signal O utput
a.
Set:
VERTICAL MODE
CH 1 and CH 2
CH 3 and CH 4
BW LIMIT
VOLTS/DIV
CH 1 and CH 2
Input Coupling
CH 1 and CH 2
A and B SEC/DIV
TRIGGER MODE
SOURCE
COUPLING
b.
Push the CH 2 VERTICAL MODE button so that light is off.
c.
Connect the CH 2 signal from the rear-panel CH 2 SIGNAL OUT con­
nector to the CH 1 OR X input connector via a 50 f l BNC cable.
d.
Align the CH 1 trace to the center graticule line.
e.
Set CH 1 Input Coupling to DC.
f.
CHECK—Display amplitude to 4.5 to 5.5 divisions (neglect trace
width).
g.
Connect a 1 kHz, 10 mV standard-amplitude signal from the Calibra­
tion Generator to the CH 2 Input Connector via a 50 f l BNC cable.
7-50

OPTION 15

Two 50 f l BNC coaxial cables
50 f l Precision BNC coaxial cable
Two 50 f l BNC termination
On (light on)
Off (light off)
Off (light off)
2 MV
GND
1 ms
AUTO LVL
VERT
NOISE REJ
2245A Operators