Performance Check Procedure
17. CH 3 and CH 4 V e rtical Bandw idth
a. Set:
VERTICAL MODE
CH 3 and CH 4 VOLTS/DIV
b.
Connect leveled sine-wave generator (SG 503) output to the CH 3 in­
put via a 50 n precision BNC coaxial cable and a 50 f l BNC termina­
tion.
c.
Set the generator output for a six-division signal display at 50 kHz.
d.
Set the generator Frequency Range and Frequency Variable controls
for a 90 MHz output frequency.
e.
CHECK—that the signal display amplitude is 4.2 divisions or more as
the frequency is increased to 100 MHz.
f.
Repeat the procedure for 0.5 VOLTS/DIV setting.
g.
Move the test signal to the CH 4 input.
h.
Set VERTICAL MODE to CH 4
i.
Repeat the procedure for CH 4.

18. SCOPE BW (Bandwidth Lim it) A ccu ra cy

a.
Set:
VERTICAL MODE
CH 1 VOLTS/DIV
SCOPE BW
b.
Move test signal from the CH 4 input to the CH 1 input.
c.
Set leveled sine-wave generator (SG 503) output for a six-division sig­
nal amplitude at 50 kHz.
d.
Increase the leveled sine-wave generator output frequency, using the
Frequency Range and Frequency Variable controls until a signal display
amplitude of 4.2 divisions is obtained.