Page 7-33
Replace part c of step 6 with the following:
c.
Increase leveled sine-wave generator output level until a stably trig­
gered display is just obtainable.
Page 7-49
AUTO SETUP Functional Check
Replace steps 2 and 3 with the following procedures:
2. P RO BE A D JU S T O utput
a.
Set:
Test Oscilloscope VOLTS.DIV
Test Oscilloscope SEC/DIV
b.
Connect a 10X probe to the test oscilloscope and connect the probe tip
to the 2245A PROBE ADJUST output.
c.
C H EC K-For a 0.5 V (± 2% into 1 MO) of PROBE ADJUST square-
wave signal (square-wave period is typically 1 ms, within 25%).
Remember to take into account the oscilloscope and probe speci­
fications when determining the accuracy of the PROBE ADJUST
square-wave signal.
3. A U TO S ETUP Functional Check
a.
Disconnect the 10X probe from the test oscilloscope and connect it to
the 2245A CH 1 input, leaving the probe tip attached to the PROBE
ADJUST output.
a.
Set:
CH 1 COUPLING
CH 1 VOLTS/DIV
A SEC/DIV
b.
Press the AUTO SETUP button.
c.
Check that the Probe Adjust waveform is stably displayed on the upper
half of the crt.
d.
Disconnect the test setup.
m a n u a l c h a n g e in f o r m a t io n
Date: 1-14-91 Change Reference:
D E S C R IP T IO N
Step 6. Trigger Level Control Range
Step 2. PROBE ADJUST Output and Step 3.
Page 1 of 1
10 mV
0.2 ms
NOTE
GND
2 mV
20 ns
C4/0190 (REV2)
Product Group 46