Notices On Emc - Atmos Record 55 Operating Instructions Manual

With standard rail
Table of Contents

Advertisement

12 Notices on EMC

 Medical electrical equipment is subject to special precautions with regard to EMC
and must be installed acc. to following EMC notes.
Guidance and manufacturer's declaration - ambient conditions
The ATMOS Record 55 is suitable for use in the following environments:
• In professional health care facilities, such as: Medical practices, clinics, first aid
facilities and operating theaters. Not suitable is the environment of HF surgical
devices and outside of an RF-shielded room of a magnetic resonance imaging
system.
• Special environments such as factory or military facilities and medical areas near HF
surgical devices, short wave therapy equipment or within a RF shielded room of a
magnetic resonance imaging system.
The customer or user of the ATMOS Record 55 must ensure that it is used in such an
environment.
Guidance and manufacturer's declaration - key features
 Please note the technical data in this manual. The essential features are fully usable
even in the presence of electromagnetic disturbances.
Guidance and manufacturer's declaration - for accessories, transducers and cables
The ATMOS Record 55 has the following electrical components:
Type
Power cable
Guidance and manufacturer's declaration - Warnings
WARNING
The use of third-party accessories, transducers and cables as specified or provided
by the manufacturer may result in increased electromagnetic emissions or reduced
immunity to electromagnetic interference and result in erroneous operation.
WARNING
Portable RF communications equipment (for example, radios, antenna cables) should
not be used within 30 cm of * the manufacturer's designated parts or lines of the
ATMOS Record 55. Failure to do so may result in a reduction in the performance of the
device.
 * At higher immunity test levels the distance may be reduced.
WARNING
Placement on or next to another device should be avoided. This could result in incorrect
operation. If this is unavoidable, the proper functioning of the device must be monitored
regularly. Please switch off any nearby devices which are not in use, if possible.
REF
Max. cable length
008.0629.0
5 m
Notices on EMC