Hoses - Atmos Record 55 Operating Instructions Manual

With standard rail
Hide thumbs Also See for Record 55:
Table of Contents

Advertisement

5.4.4 Hoses

Pretreatment at the site of use
Collect and transport
Pre-cleaning
Dismantle
Mechanical cleaning and disinfection
Pre-wash 1 min
Cleaning: 5 min, 55°C / 131°F
Neutralize 2 min
Intermediate rinse 1 min
Disinfect 5 min, 93°C / 199°F
Dry: 12 min, 110°C / 230°F
Check and maintain
Assembly
Function check
Packaging
Sterilize
Prefractionated vacuum: 3x
Temperature: 134°C / 273°F
Time: 5 min
Dry: 10 min
Storage
• Clean the accessories under cold, running water.
• Thoroughly rinse the cavities of the accessories with running
water.
There is no more residue visible.
• Label any damaged accessories.
• Place the accessories in a rinsing canister.
• Close the rinsing canister.
• Transport the rinsing canister to the treatment site.
• Clean the accessories under running water.
• Thoroughly rinse the lumen of the accessories with running
water.
Not necessary.
• Secure the accessories on a suitable load carrier.
• Clean and disinfect using a suitable program:
– Pre-rinse with cold water
– Cleaning with cleaning agent
– Neutralize with cold water
– Intermediate rinse with softened, cold water
– Disinfection with demineralized water
– Drying
• Cleaning and disinfection device: according to EN ISO
15883-1
• Adapter: Miele E366/E446
• Program: Miele Vario TD
• Check the success of the reprocessing with a suitable light
magnifier.
• If reprocessing was unsuccessful, the procedure must be
repeated.
• Dispose of damaged accessories or have them repaired.
Not necessary.
Not necessary.
• Label damaged accessories.
• Pack the accessories with a packing system according to DIN
EN ISO 11607.
• Sterilize the accessories using a suitable procedure:
– Steam sterilization / autoclaving
Sterilisator: according to EN 285.
• Observe the ambient conditions, see Chapter "11 Technical
data".
Reprocessing
23