Pretreatment at the site of use
• Flushing: 60 s
• Rinsing: 60 s
Collect and transport
Dismantle
Pre-cleaning
• Flushing: 1x / 30s
• Rinsing: 60 s
Brush: Round brush
Size: 7 mm, Material: Nylon
Brush: Round brush
Size: 11 mm, Material: Nylon
Brush: Round brush
Size: 15 mm, Material: Nylon
Brush: Square
Size: 40 x 10 mm, Material: Nylon, Special features: with
angled head
Mechanical cleaning and disinfection
Pre-wash 1 min
Cleaning: 5 min, 50°C / 122°F
Neutralize 2 min
Intermediate rinse 1 min
Disinfect 5 min, 93°C / 199°F
Dry: 12 min, 110°C / 230°F
Check and maintain
Assembly
Function check
Packaging
Sterilize
Prefractionated vacuum: 3x
Temperature: 134°C / 273°F
Time: 5 min
Dry: 10 min
Storage
22 Reprocessing
• Empty the rinsing canister
• Clean the accessories under cold, running water.
• Thoroughly rinse the cavities and lumens of the accessories
with running water.
There is no more residue visible.
• Label any damaged accessories.
• Place the accessories in a rinsing canister.
• Transport the rinsing canister to the treatment site.
• See chapter "5.2 Preparing / completing reprocessing"
• Dispose of disposable products
• Make the following cavities accessible:
– Double hose connector
– Canister lid
• Make the following lumens accessible:
– Double hose connector
• Thoroughly clean the accessories evenly with a suitable
brush under running water
• Thoroughly rinse the cavities and lumens of the accessories
with running water.
• Secure the accessories on a suitable load carrier.
• Clean and disinfect using a suitable program:
– Pre-rinse with cold water
– Cleaning with cleaning agent
– Neutralize with neutralizing agent
– Intermediate rinse with softened, cold water
– Disinfection with demineralized water
– Drying
• Cleaning and disinfection device: according to EN ISO
15883-1
• Program: Miele Vario TD
• Check the success of the reprocessing with a suitable
light magnifier. They must be free of particles and organic
material.
• If reprocessing was unsuccessful, the procedure must be
repeated.
• Dispose of damaged accessories or have them repaired.
• Not necessary.
• Not necessary.
• Label damaged accessories.
• Pack the accessories with a packing system according to DIN
EN ISO 11607.
• Sterilize the accessories using a suitable procedure:
– Steam sterilization / autoclaving
• Sterilizer: According to DIN EN 285
• Observe the ambient conditions, see Chapter "11 Technical
data".