Unsuitable cleaning agents and disinfectants.
Risk of damage to the product.
•
Do not use any process chemicals which contain the following ingredients on plastic
parts
• Chloramide or phenol derivatives
•
Do not use any process chemicals which contain the following ingredients on
stainless steel
• Organic or inorganic bases
• Alkaline solutions
Incorrect mechanical cleaning and disinfection.
Corrosion due to moisture.
•
When the program is finished remove the products immediately.
5.2 Preparing/completing reprocessing
Prior to reprocessing
1. Disassemble the product for reprocessing into the following items:
• Device
• Hoses
• Secretion canister system
After reprocessing
1. Perform a function check.
5.3 Prepare surfaces
5.3.1 Overview
Surface
Housing
x
x
Remarks
x
Reprocessing
19