First start-up ......................13 ....................13 reParing the device Connecting the device to the mains power supply ..........13 Connecting the DDS secretion canister (ATMOS Twin Record 55 DDS) .... 14 4.4.1 Assembly of the DDS secretion canister ............14 4.4.2 Using the DDS splash protection ..............
Page 3
5.4.2 Selecting process chemicals ................25 5.4.3 Secretion canister system ................. 26 5.4.4 Hoses ........................27 Reprocessing the accessories (ATMOS Twin Record 55 with standard rail) ..28 5.5.1 Overview ......................28 5.5.2 Selecting process chemicals ................29 5.5.3 Secretion canister system ................. 29 5.5.4...
The quality management system at ATMOS has been certified according to international standard EN ISO 13485. These operating instructions are valid for the following devices: ATMOS Twin Record 55 DDS REF 443.0950.0 ATMOS Twin Record 55 with standard rail REF 443.0960.0 Introduction...
1.2 Explanation of pictures and symbols In the operating instructions DANGER Warning of a danger that will result in immediate fatal or serious injury. Observe the necessary measures. WARNING Warning of a danger that can cause fatal or serious injury. Observe the necessary measures.
European Article Number Protection against the penetration of harmful moisture (dripping water) Applied part type BF Professional disposal For single use only (symbol located on consumables) Not sterile STERILE Autoclavable AUTOCLAVE PATIENT Connection for suction hose / patient Potential equalisation Protection class II Fuse Alternating current...
1.4 Function The ATMOS Twin Record 55 is a mains-operated surgical suction device in which two ATMOS Record 55 are combined in one shell. The core piece are two high-performance, maintenance-free diaphragm pumps. These generate vacuum in the suction hose and secretion canister system, which assists in suctioning and collecting secretions. The final...
Only transport the product in a shipping carton that is padded and offers sufficient protection. If damage occurs during transport: 1. Document and report damages in transit. 2. Send the device to ATMOS see chapter „6.3 Sending in the device“ on page 32. Ambient conditions for transport and storage: • Temperature: −30...+50 °C • Relative humidity: 5...90% without condensation...
„1.3 Intended use“ on page 6. • Never operate the product in potentially explosive areas or in areas that are oxygenated. • Only use original accessories and original spare parts from ATMOS. WARNING Your patient can be severely injured. Avoid improper use. •...
Page 10
• Remove the transport protection on the bottom of the device prior to first use. • ATMOS recommends always having another suction device ready at hand. Thus suction is available even in the event of device failure. WARNING Risk of infection due to pathogens on the product! Deadly diseases can be transmitted.
Only use proper power cables and extension cords. • Never touch the device’s interfaces and the patient at the same time! • Use only original accessories and original spare parts from ATMOS. • Please pay attention to the period tests in chapter 6.0 „Maintenance and service“. •...
Single docking station Connection for the foot switch (foot controller for switching the left pump ON or OFF) Front view ATMOS Twin Record 55 with standard rail Vacuum gauge I Vacuum regulator I ...
2. Check the device for any transport damage. 3. If the device is damaged: Document and report the transport damage. Send in the device to ATMOS „6.3 Sending in the device“ on page 32 4.3 Connecting the device to the mains power supply A medical isolation transformer according to EN 60601-1 is required if several devices are connected over one common power supply.
4.4.5 Attaching, closing and opening the DDS secretion canister handle 1. To attach the DDS secretion canister handle, insert it into the grooves of the canister lid (with the locking latches open). 2. To close the DDS secretion canister handle, clip the locking latches under the canister rim.
4.4.10 Suction 1. Please ensure that the following parts have been reprocessed before treating a new patient: • Suction hose including hose-rinsing aperture or suction instruments • DDS secretion canister system including DDS canister lid and DDS hose adapter 2. Prior to each use, check whether the DDS bacterial filter was inserted during cleaning and disinfection.
5. If the main switch in the control panel is set to continuous operation (ON), the foot switch produces no effect. 4.5 Connecting the secretion canister (ATMOS Twin Record 55 with standard rail) 4.5.1 Connecting the bacterial filter • Insert the bacterial filter with the short hose onto the connection piece on the device.
4.5.4 Connecting the suction hose • Attach the suction hose to the angled connection. 4.5.5 Serres®, Medi-Vac® secretion canister system, other canister systems Observe the operating instructions by the manufacturer of the corresponding canister system. Do not operate the device without a bacterial filter. 4.5.6 Secretion canister system •...
4.5.8 Using an overflow canister If the bacterial filter becomes frequently blocked, ATMOS recommends integrating an additional overflow canister (REF 444.0646.0) between the secretion canister and the bacterial filter. This absorbs moisture and foam bubbles occurring. An additional DDS bacterial filter can be used in the overflow canister. Replace the bacterial filter (REF 340.0054.0) with a new filter if it is discoloured, contaminated or oversucked. When the filter is used regularly, a filter change is recommended after 14 days at the latest. 4.5.9 Using a smoke evacuation filter According to the intended use, the device is used to suction liquids and pieces of tissue.
4.5.11 Checking the bacterial filter ATTENTION The bacterial filter is a disposable product and cannot be autoclaved or disinfected. • Set the vacuum regulator to "max". • As soon as the vacuum gauge shows a vacuum value highter than -0.3 bar, while the suction hose is open, the filter must be replaced. • To do this, remove the suction connections on the bacterial filter and insert the new bacterial filter. Pay attention to the correct flow direction. • Dispose of the used bacterial filter immediately. Always have some spare filters ready to hand! 4.5.12 Exchanging the secretion canister 1.
4.5.14 Options Foot switch, REF 443.0755.0 Pneumatically explosion-proof switch for switching the device on and off. 1. Connect the foot switch. 2. Set the main switch in the control panel to foot switch operation (OFF). 3. Pressing the foot switch turns the device on. 4. Pressing the foot switch again turns the device off.
5 Reprocessing 5.1 Safety instructions for reprocessing 5.1.1 General safety instructions We recommend that you always document all maintenance work and part replacements in writing. It is the responsibility of the user to ensure that the required results of cleaning and disinfection are achieved.
Unsuitable cleaning agents and disinfectants. The product can become damaged. • Do not use any process chemicals containing the following ingredients on plastic parts: • Chloramides or phenol derivatives • Do not use any process chemicals containing the following ingredients on stainless steel: •...
5.3.2 Selecting process chemicals Follow the manufacturer’s instructions for the process chemical. Cleaning agent (manufacturer) Active ingredients in 100 g Type Disinfection Green & Clean SK <1 g dialkyldimethylammonium chloride, <1 g Foam (Metasys) alkyldimethylethylbenzylammonium chloride, <1 g Ready to use alkyldimethylbenzylammonium chloride Dismozon®...
5.4.3 Secretion canister system Characteristics The accessories have the following hard-to-reach areas: • Double hose connector (lumina) • Secretion canister lid complete (hollow spaces) Take particular care when reprocessing hard-to-reach areas. Pre-treating at the 1. Empty the canister. site of use 2.
Checking and 1. Check whether reprocessing was successful with a maintaining suitable light magnifier. • Free of particles and organic material 2. Dispose of damaged accessories or have them repaired. Assembly Not necessary. Function check Not necessary. Packaging 1. Label the accessories. 2.
Temperature: 134°C / 273°F Time: 5 min Dry: 10 min Storage Observe the ambient conditions; see chapter „11 Technical data“ on page 38. 5.5 Reprocessing the accessories (ATMOS Twin Record 55 with standard rail) 5.5.1 Overview Accessory Secretion canister system •...
5.5.2 Selecting process chemicals Follow the manufacturer’s instructions for the process chemical. Cleaning agent (manufacturer) Active ingredients in 100 g Type Disinfectants – Manual reprocessing Gigasept® FF new < 5 % phosphonate, < 5 % anionic surfactant, < 5 Liquid (Schülke &...
Page 30
Pre-cleaning 1. Make the following hollow spaces accessible: – Double hose connector • Flushing: 1x / 30s – Secretion canister lid • Rinsing: 60 s 2. Make the following lumina accessible: Brush: Circular brush Size: 7 mm, Material: Nylon – Double hose connector Brush: Circular brush 3.
5.5.4 Hoses Take particular care when reprocessing hard-to-reach areas. Pre-treating at the 1. Clean the accessories under cold, running water. site of use 2. Thoroughly rinse the hollow spaces of the accessories with running water. » No more residue is visible. Collecting and 1.
ATMOS recommends: work should be carried out by an authorised ATMOS service partner. This ensures that repairs and tests are carried out professionally, original spare parts are used, and warranty claims are maintained.
7 Troubleshooting 7.1 Troubleshooting The product has been subjected to thorough quality control in the factory. However, if a fault should occur, you may be able to resolve it yourself. Error symptom Possible cause Remedy • Device does not • Power plug fits •...
8 Accessories Accessories Foot switch 443.0755.0 Foot controller set for the ATMOS Twin Record 55 443.0770.0 Potential equalisation cable 008.0596.0 Practice package 1.5 l 340.0002.0 Practice package 3 l 340.0003.0 DDS secretion canister, plastic 1.5 l, autoclavable 340.0050.0 DDS secretion canister, plastic 3 l, autoclavable 340.0051.0...
Page 35
Accessories Receptal® holder complete for standard rail 1 x 3 l 444.0170.0 Receptal® holder complete for standard rail (single) 444.0150.0 Receptal® holder complete for standard rail (double) 444.0156.0 Holder for Receptal® canister, with equipment mount hM57525661 Receptal® package 1 x 2 l for mounting to a standard rail 444.0030.0 Receptal®...
9 Consumables Spare parts Bacterial filter for the ATMOS DDS secretion canister, pack of 10 pcs. 340.0054.0 Suction hose, PVC, disposable, Ø 8 mm, L = 2.10 m, 50 pcs. 006.0059.0 Suction hose, silicone, Ø 6 mm, L = 1.30 m, 1 pc. 000.0013.0 Suction hose, silicone, Ø 6 mm, L = 2 m, 1 pc.
The product does not contain any hazardous materials. 1. Clean and disinfect the product. 2. In Germany: send the product back to ATMOS or your specialist dealer. They will dispose of the product properly. 3. For other countries: Dispose of the product properly and in accordance with country-specific laws and regulations.
11 Technical data 11.1 ATMOS Twin Record 55 DDS Voltage 230 V~ ± 10%; 50/60 Hz Current consumption for each pump approx. 0.45 A (230 V~) Power consumption Approx. 200 W Fuses 2x T 1.25 A/H (250 V~) Temperature protection switch 150°C...
* 1 bar ≌ 750,06 mm Hg ≌ 1000 hPa / dependent on daily air pressure ID No. (REF) 443.0950.0 atMOS Twin Record 55 ddS 11.2 ATMOS Twin Record 55 with standard rail Voltage 230 v~ ± 10%; 50/60 h Current consumption for each pump approx.
Page 40
Secretion canister 444.0034.0 Graduated secretion jar 5 l 444.0100.0 Secretion canister 1.5 l (PC) 444.0099.0 Secretion canister 3 l (PC) HM57525656 Secretion canister 4 l PSU with equipment mount HM57525658 Secretion canister 4 l PC with equipment mount 310.0402 x Serres®outer canister 2 l 310.0403.0 Serres®-outer canister 3 l 443.0256.0 Receptal®...
Page 41
63642 (Surgical Suction Pump) cOde UMdnS 10-217 (Aspirators, surgical) cOde Md/MdS Md 1104 (Active surgical devices)/ cOde Mda 0312 (Other active non-implantable surgical device) 42503651SurgicalRecord5Jv aSic device identifier ID No. (ref) 443.0960.0 atMOS Twin Record 55 with standard rail Technical data...
HF-shielded room of a magnetic resonance imaging system. The customer or user of the ATMOS Twin Record 55 must ensure that it is used in such an environment. Guidance and manufacturer’s declaration – key features ...