7 Troubleshooting

7.1 Troubleshooting
The product has been subjected to a thorough quality control in the factory. However if
a fault should occur, you may be able to resolve it yourself.
Error symptom
• Device does not
start
• Not enough
power
• No suction
capacity
Possible cause
• Power plug is fitted
badly
• No mains voltage
• Defective fuse
• Leakage in the hose
or in the secretion
canister system
• Bacterial filter is
blocked (vacuum
gauge indicates
vacuum)
• The float of the
overflow protection
seals the double hose
connector
• Secretion or blood
was sucked in and
the valve plates of
the aggregate are
stuck together
Remedy
• Check the connection to the
socket and the device
• Check main fuse
• Exchange the fuse
• Check the canister lid and hoses
for tightness, replace sealing ring
on the canister lid if necessary
• Replace the bacterial filter
• Check fluid level in the secretion
canister, If necessary it must
be emptied. Clean overflow
protection and check float for
free movement
• In this case, the device must be
sent in for repair
Troubleshooting
25