Function; Intended Users - Atmos Record 55 Operating Instructions Manual

With standard rail
Hide thumbs Also See for Record 55:
Table of Contents

Advertisement

Specification of the
main function:
User profile:
Patient groups:
Application organ:
Application time:
Area of application:
Contraindications:
The product is:
Sterility:
Single-use product/
re-sterilisation:

1.4 Function

The ATMOS Record 55 is a mains-operated surgical suction device, the core of which is a
high-performance, maintenance-free diaphragm pump. It generates vacuum in the hose
and rinsing canister system which assists in drawing off and collecting the secretions.
Via a vacuum regulator with vacuum gauge, the final vacuum and thus the desired
suction capacity can be precisely adjusted.
Several secretion canisters of different sizes are available for secretion collection. A
mechanical overflow protection on the double hose connector and a bacterial filter
in the vacuum hose from the secretion canister to the pump, prevents the suction of
secretion into the pump.

1.5 Intended users

May only be used by trained professionals in supervised and medical operations.
For the drainage and temporary collection of body fluids. By
means of an electrical suction pump a negative pressure is
created. An additional secretion canister must be attached to
allow temporary collection of the derived body fluids.
Doctors, medical auxiliaries without restrictions.
Patients of all ages with and without restrictions.
Natural orifices and openings which result from a surgical
intervention (entire body of humans and animals).
For short term use on patients (< 30 days)
The application site is the clinical, outpatient and veterinary
field. The device may only be used by persons who have
received the relevant training and instruction.
Not suitable for:
• Drainage operations in the low vacuum range (e.g.
thoracic or wound drainage)
• Use outside the medical sector
• Suction of flammable, corrosive or explosive substances
• Suction in potentially explosive atmospheres
• Not suitable for use as a vacuum extraction system
Active
No sterile product
The device and parts of the accessories are reusable. For
information on reprocessing, cleaning and disinfection please
see the operating instructions.