6.0 Cleaning

6.1 General information on cleaning and disinfection

Prior to cleaning
Medical devices must always offer a maximum in safety and function.
We therefore recommend:
Prior to each application:
The measures described herein regarding cleaning
and disinfection do not replace the regulations
valid for operating the device!
The information regarding concentration of
cleaning liquids and disinfectant solutions stated by
the relevant manufacturers must strictly be
observed!
if
necessary
6.2
Cleaning and sterilizing hoses and
collection jar
Before using the device on a new patient be sure
to clean and sterilize the following parts:
– suction hose including suction attachment or aspi-
ration set,
– DDS collection jar including DDS collection jar
lid, DDS hose adapter, splash protector and DDS
collection jar handle.
Unscrew all hose connectors, pull the DDS hose adap-
ter out of the DDS collection jar lid, open the lid, empty
the collection jar and dispose of the suctioned material
properly.
Take the DDS bacterial fi lter out of the DDS collection jar
handle.
All other parts, except the bacterial fi lter, must also be
thoroughly rinsed under running water. Using the cleaning
agent Neodisher AN (manufactured by Dr. Weigert, Ham-
burg) cleaning in an automatic cleaner and disinfecter is
also possible.
Thermal disinfection is carried out at 93° C.
A u t o c l a v e a l l o f t h e p a r t s r e f e r r e d t o a b o v e
(up to 134 °C, 3 bar, 20 min max).
After sterilization, reassemble all parts
(section 4.0 "Operation").
For disinfection, you may use all surface and instrument
disinfectants listed in chapter 6.4 / 6.5.