FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | CI/FEP630/FEH630-EN REV. D
Zero point adjustment of the flowmeter
Note
Prior to starting the zero point adjustment, make sure that:
There is no flow through the sensor (close all valves, shut-
off devices etc.)
The sensor is completely filled with the medium to be
measured
Easy Setup
System Zero
Next
Use
to start automatic adjustment of the zero point for
the system.
Menu
Easy Setup
Exit
Select
Once all parameter have been set, the main menu appears again.
The most important parameters are now set.
40. Use
to switch to the process display.
Change from two to one column
OK
53