FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | CI/FEP630/FEH630-EN REV. D
Menu
Easy Setup
Exit
Select
5. Use
/
to select 'Easy Setup'.
6. Confirm the selection with
Easy Setup
Language
Language
Next
7. Use
to call up the edit mode.
8. Use
/
to select the desired language.
9. Confirm the selection with
Easy Setup
Unit Volumeflow Qv
Next
10. Use
to call up the edit mode.
11. Use
/
to select the desired unit for the volume flow
rate.
12. Confirm the selection with
.
Edit
.
l/s
Edit
.
Easy Setup
Qv Max 1
25.000 l/s
Next
13. Use
to call up the edit mode.
14. Use
/
to set the desired upper range value.
15. Confirm the selection with
The device is factory calibrated to the flow range end value
Q
DN, unless other customer information is available. The
max
ideal upper range values are those which correspond to a flow
velocity of 2 to 3 m/s (0.2 to 0.3 × Q
The adjustable upper range values are listed in the table at
Measuring range table on page 54.
Easy Setup
Unit Vol. Totalizer
Next
16. Use
to call up the edit mode.
17. Use
/
to select the desired unit for the volume
totalizer.
18. Confirm the selection with
Easy Setup
Pulses per Unit
10.000/m³
Next
19. Use
to call up the edit mode.
20. Use
/
to select the desired pulse per unit for the
pulse output.
21. Confirm the selection with
Edit
.
DN)
max
l/s
Edit
.
Edit
.
51