FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | CI/FEP630/FEH630-EN REV. D
Flowmeter sensor in high temperature version
Lining material
Flange material
Thick PTFE***
Steel
Thick PTFE***
Stainless steel
PFA***
Steel
PFA***
Stainless steel
ETFE***
Steel
ETFE***
Stainless steel
*
Only for China production site
** For (optional) low-temperature version only
*** Only for design level 'A'
***
Ambient temperature range (T
Minimum
-10 °C (14 °F)
-20 °C (-4 °F)
40 °C (-40 °F)**
-10 °C (14 °F)
-20 °C (-4 °F)
40 °C (-40 °F)**
-10 °C (14 °F)
-20 °C (-4 °F)
40 °C ( 40 °F)**
)
Measuring medium temperature (T
amb
Maximum
Minimum
60 °C (140 °F)
-10 °C (14 °F)
60 °C (140 °F)
-20 °C (-4 °F)
60 °C (140 °F)
-10 °C (14 °F)
60 °C (140 °F)
-20 °C (-4 °F)
60 °C (140 °F)
-10 °C (14 °F)
60 °C (140 °F)
-20 °C (-4 °F)
71
)
medium
Maximum
180 °C (356 °F)
180 °C (356 °F)
180 °C (356 °F)
180 °C (356 °F)
130 °C (266 °F)
130 °C (266 °F)