8588A/8558A
Remote Programmer's Manual
:RANGe? [MIN|MAX|DEF]
:RESolution <nrf>|MIN|MAX|DEF
:RESolution? [MIN|MAX|DEF]
[:SENSe]:XCURrent:AC
:COUNter:BLIMit[:STATe] <boolean>
:COUNter:BLIMit[:STATe]?
:COUNter:COUPling AC | DC
:COUNter:COUPling?
:COUNter:GATE:AUTO[:STATe]<boolean>
:COUNter:GATE:AUTO[:STATe]?
:COUNter:GATE < nrf >|MIN|MAX|DEF
:COUNter:GATE? [MIN|MAX|DEF]
:COUPling AC|DC
:COUPling?
:FILTer <nrf>|MIN|MAX|DEF
:FILTer? [MIN|MAX|DEF]
36
Table 6. Command Summary (cont.)
Keyword
Description
Returns the selected range, or if specified, the MIN,
MAX, or Default range
Set maximum expected value or min, max or default
resolution; for example, range is 1V, Resolution
<nrf>. = 0.0001 (100 µV), the measurement is
returned with a resolution of +1.000E-4
Returns the selected resolution, or if specified, the
minimum, maximum or default range
Turn the frequency path bandwidth limit ON or OFF
Return the state of the frequency path bandwidth limit
setting. Returns 1 for Auto range ON, 0 for auto
range OFF.
Set the frequency path coupling to ac or dc. Note that
if the signal path coupling is ac, frequency path
coupling is forced to ac.
Return the frequency path coupling setting
Turn Auto selection of counter gate ON or OFF
Return the state of the counter gate auto setting, 1 for
Auto gate ON, 0 for auto gate OFF
Sets the secondary frequency counter measurement
gate
Return counter gate setting, or if specified, the MIN,
MAX, or Default setting
Set the ac current signal path coupling to ac or dc.
Note that if the signal path coupling is ac, frequency
path coupling is forced to ac.
Return the frequency path coupling setting
Sets the input filter to value or to the minimum,
maximum or default setting. The filter setting range is
0.1Hz to 1kHz, the discrete settings are: 0.1 Hz,
1 Hz, 10 Hz, 40 Hz, 100 Hz, 1 kHz. The discrete
setting closest to <nrf> will be selected.
Return the filter setting