Prepare The Product For Ethernet Remote Control - Fluke Calibration 8588A Remote Programmer's Manual

Reference multimeter and 8 1/2 digit multimeter
Table of Contents

Advertisement

To ensure the most efficient use of the Service Request (SRQ)
mechanism provided by USBTMC-USB488, first check that the RQS
bit in the Status Byte is clear. To do this, perform a serial poll of the
Product before you send any commands that could generate SRQ
events. This only needs to be done once in your program/procedure.
Check communications can be established:
Write *IDN?
Read the response
Clear registers and read the status byte:
Write: *CLS;*RST
Write *ESE 0
Write *SRE 191
Read the status byte (Serial POLL)
Before you proceed, set the event and status registers as required.

Prepare the Product for Ethernet Remote Control

To prepare the Product for Ethernet remote control?
1. Push  and select Remote Settings.
2. Use  and  to highlight the Active Port row.
3. Select  (Ethernet).
4. Use  and  to highlight Ethernet: Settings and push .
The following items can now be configured:
DHCP
Set to  (OFF) for a direct Product to PC connection or if a fixed IP
address is required.
Set to ON to allow automatic allocation of the IP address by a DHCP
server.
Ethernet IP Address
Unique address for the Product. This will be automatically allocated if
DHCP is set to ON otherwise this must be manually set if DHCP is OFF.
The default is 127.000.000.001 which is the loopback internet protocol or
localhost which cannot be used for external connections.
Gateway
Default: 192.168.001.001
Reference Multimeter and 8 ½ Digit Multimeter
Prepare the Product for Ethernet Remote Control
Note
7