Keyword
CALibration
CALibration
:COPY:BASeline
:SECure?
:SECure:EXIT
:SECure:PASSword <string>
:STORe:DATe? <CERTified|BASeline>
:STORe:USE <CERTified|BASeline>
:STORe:USE?
:TARget:DESCription?
:TARGet:INPut?
:TARGet:STEP <nrf>
:TARGet:VALue <nrf>
:TARGet:VALue[?]
:TRIGger?
DISPlay Subsystem
DISPlay[:STATe]<boolean>
DISPlay
:BRIGhtness <nrf>
:BRIGhtness?
:DTIMe<nrf>
:DTIMe?
DISPlay[:STATe]?
Table 6. Command Summary (cont.)
Copy the certified stores over the Baseline stores
Returns calibration adjustment mode "FACTory", "USER" or
"NONe"
Exit from calibration adjustment mode
Enables calibration adjustment mode
Returns the date of the last calibration adjustment of the store
<year>,<month>,<day>,<Hours>,<Minutes>,<seconds>
Sets the active stores. Default CERTified.
Returns the active stores "CERTified" or "BASeline"
Returns the target name as a quoted string containing target
point description
Returns the target input field description as a quoted string
Sets the calibration adjust step
Sets the target value
Returns the target value
Initiates the adjustment, returns 0 = success, 1 = failure
Turn the display ON or OFF. Default ON.
Set the display brightness. Default 50 %
:BRIGhtness? Retrieve Brightness setting
Set the display backlight timer in seconds. Default 1800
seconds (30 minutes)
Returns the time setting
Returns the state of the display
Reference Multimeter and 8 ½ Digit Multimeter
Description
SCPI Commands Reference
21