8588A/8558A
Remote Programmer's Manual
[:SENSe]:FRESistance
:APERture <nrf>|MIN|MAX|DEF
:APERture? [MIN|MAX|DEF]
:APERture:MODE AUTO|FAST|MANual
:APERture:MODE?
:NPLC <nrf>MIN|MAX|DEF
:NPLC? [MIN|MAX|DEF]
:MODE NORMal|TRUE|HIV
:MODE?
:LOWI[:STATe] <boolean>
:LOWI?
:RANGe:AUTO[:STATe] <boolean>
:RANGe:AUTO[:STATe] ?
:RANGe <range>|MIN|MAX|DEF
:RANGe? [MIN|MAX|DEF]
:RESolution <resolution>|MIN|MAX|DEF
:RESolution? [MIN|MAX|DEF]
:FILTer[:STATe] <boolean>
:FILTer[:STATe]?
[:SENSe]:FUNCtion <string>
(where string can be)
"CURRent[:DC]"
"CURRent:AC"
"CAPacitance"
28
Table 6. Command Summary (cont.)
Keyword
Description
Sets the ADC aperture value in seconds or to the MIN, MAX, or
default setting
Returns the aperture setting or the MIN, MAX, or Default
settings in seconds
Sets the aperture mode
Returns the aperture mode
Sets the ADC aperture in number of power line cycles or to the
MIN, MAX, or Default plc setting
Returns the ADC aperture in number of power line cycles or to
the MIN, MAX, or Default plc setting
Sets the 4-wire Ohms mode
Returns the 4-wire Ohms mode setting
Sets the Low current mode ON or OFF
Returns the low current mode setting
Sets Auto ranging ON or OFF
Returns the Auto range setting
Sets the expected maximum input resistance, or if selected, to
the MIN, MAX, or Default setting
Returns the range setting, or if selected, the MIN, MAX, or
Default setting
Set maximum expected value or min, max or default resolution
Returns the selected resolution, or if specified, the MIN, MAX, or
Default resolution
Turns the input filter ON or OFF
Returns the filter setting, 0 = OFF, 1 = ON
Sets function to DCI
Sets function to ACI
Sets function to Capacitance
Need help?
Do you have a question about the 8588A and is the answer not in the manual?