Sampler Correction Constants Routine - HP 8753E Service Manual

Network analyzer
Table of Contents

Advertisement

minutes.
If the analyzer displays Sampler Cor - FAIL, check the following:
n
a. The HP-IB address of your power meter is set at 13. Then rerun this
routine ("Sampler Correction Constants Routine").
b. The HP 8482A power sensor is connected to Port 1. Rerun this routine

('Sampler Correction Constants Routine ").

19. For Option 006 Instruments Only: When the analyzer displays CONNECT
6 GHz SENSOR B TO PORT 1, make the connections as shown in Figure 3-6.
Then press @WJ~. This part of the test will take about 20 seconds.
Figure 3-6. Connections for Sampler Correction at 6 GHz
20. When the analyzer displays CONNECT <3 GHz SENSOR A TO PORT 2, make the
connections as shown in Figure 3-7.
Adjustments and Correction Constants