Power Sensor Calibration Factor Entry
5. Press @--+q ~~~~~~~~ ~~~~~~~~~~ ~~~~~~~~~~..;~~~~3..
sensor A (HP 8482A for a 500 analyzer, or HP 8483A Option HO3 for a 75Q
analyzer).
6. Zero and calibrate the power meter and power sensor.
7. Build a table of up to 55 points (55 frequencies with their calibration
factors). 'lb enter each point, follow these steps:
b. Input a frequency value and then press the appropriate key (@J, m,
or @I).
corresponds to the frequency you entered.
The cal factor and frequency values are found on the back of the sensor.
If you make a mistake, press a and re-enter the correct value.
Note
The following terms are part of the sensor calibration menu:
Adjustments and Correction Constants
allows you to select a frequency point.
allows you to edit or change a previously entered
value.
allows you to delete a point from the sensor cal
factor table.
allows you to add a point into the sensor cal
factor table.
allows you to erase the entire sensor cal factor
table.
allows you to complete the points entry of the
sensor cal factor table.