8. Set the HP 8496A to 10 dB.
9. Set the following analyzer parameters:
10. Set up the HP 8753E for power meter calibration:
a. Select the HP 8753E as the system controller:
(Local.. _
b. Set the power meter's address:
the correct model number is displayed (HP 436A or HP 438A/437).
d. Select the cal kit and enter the power sensor calibration data.
11. 'I&e a power meter calibration sweep.
12. Verify that the power meter reads approximately -20 dBm.
Performance Tests
. . .