6. Set up the
network
. . . .
a. Press ~$Bj$gi$.
b. Enter the fundamental frequency. For example, press lso) m to enter
the hrst fundamental frequency in the 3 GHz test record.
7. Measure and record the power in the second or third harmonic by taking a
single sweep with the
a. Press @iTZVKQ.
b. Read the MARKER A measurement, and record it in the appropriate row
of the test record under Measurement Value (dBc).
8. Reset the
9. Repeat steps 5 through 8 for the remaining second and third harmonic
frequencies, and the fundamental frequencies listed in the test record.
spectrum aruzlgzer:
System Verification and
Performance Tests
2-111