Procedure; Correction Data Files - HP 84125C User Manual

Microwave emi measurement test system
Table of Contents

Advertisement

CAUTION

Procedure

3-12 System Setup and Functional Tests
your system were loaded into the spectrum analyzer at the factory or
Hewlett-Packard Service Center.
To view the test set gain and antenna correction factor data
separately, you must use the HP 84125C utility disk provided with the
HP 84125C system. The following les can be found on the utility
disk:
The test set gain correction factor les are for viewing purposes only.
not
Since these les are
made to these correction factors will result in an uncalibrated system.
Therefore a system calibration would need to be performed to ensure
accurate measurements.
Table 3-4. Correction Data Files
File Name Frequency Range
e3115.txt
e3160-09.txt 18 to 26.5
e3160-10.txt 26.5 to 40
lter 0.txt
lter 1.txt
lter 2.txt
lter 3.txt
lter 4.txt
lter 5.txt
1. Insure that the spectrum analyzer meets the conditions listed in \
HP 8564E Spectrum Analyzer Speci cations and Characteristics"
in Chapter 2.
PRESET
2. Press
on the synthesized sweeper then set the controls as
4
5
follows:
CW
Power Level
write protected, any changes inadvertently
(GHz)
Antenna Correction Factors
1 to 18
Test Set Gain
1 to 18
Band 0
1.5 to 18
Band 1
3.5 to 18
Band 2
8.25 to 18
Band 3
18 to 26.5
Band 4
26.5 to 40
Band 5
1 GHz (or one of the other Test Set
Frequencies in Table 3-5)
60 dBm